Schematic of the 18 kW setup
Specifications:
- 18 kW Ultrex Rigaku Rotating Anode (donated by Abbott Labs in 2006)
-
Cu Anode: Cu Kα1 λ = 1.5406 Å,
Cu Kα2 λ = 1.54443 Å
-
Point Source: Normal Focus 0.5mm x 10 mm, 18 kW max power
-
The effective source size should be 0.5-mm high by 1-mm-wide due to 6° take-off
angle. But we measure the width to be 0.5 mm wide (see figure).
-
Nominal operating power: 50 kV, 240 mA
- Huber slits for controlling horizontal and vertical divergence (see figure).
- Nominal setting: hsdgap = 0.7 mm wide by 14-mm-high
-
Graphite (002) Grade ZYA (< 0.5° mosaic) from Advanced Ceramics Corp.
-
Size: 25 x 25 x 2 mm
-
Para- focusing in the horizontal direction
-
Sagital-focusing in the vertical direction with Rs = 225 mm
-
2d002 = 6.708 Å, Cu Kα1 qB = 13.28°
-
1 to 1 focusing, F1 = F2 = F = RS/sin(q) = 980 mm.
This is the distance from the source to the mono
and from the mono to the sample.
-
Incident beam intensity monitor (NaI detector collecting scattered x-rays from Co coated
Kapton foil) I0 = 8900 cps at hsdgap = 0.7mm and vsdgap = 14 mm.
-
Huber incident beam slits for passing Cu Kα1 and blocking Cu
Kα2
-
Nominal setting ka2cen = 0, ka2gap = 0.5 mm-wide by 1-mm high
-
Huber 4-circle diffractometer
-
Motorized Z-sample, zsamp=0
should put pin in COR (Center of Rotation)
-
Huber 1006
X-Y- c-f goniometer
-
Sample holder
-
2-theta arm: (Nominal
settings)
-
guard slit (
gshhcen=.64,
gshgap=1.1, gsvcen=1.4, gsvgap=8
)
-
XIA filter box, ( T = 0.24 x 0.01 for
STB straight thru beam)
-
detector slit, (
dshhcen=.05,
dshgap=1.0, dsvcen=1.9, dsvgap=12
)
-
Cyberstar NaI X-ray detector
( 0 .3 m sec, 740V, Gain = 8,
LL=2.5, UL=6.5)
Auxillary equipment:
-
Laser Level on Tripod
-
X-ray Eye
-
Goniometer pin
-
High-power zoom lens camera
|